FDA cybersecurity and IEC 62304

FDA Cybersecurity for a Medical Device Cloud: What You Inherit and What You Own

What section 524B and FDA's premarket cybersecurity guidance expect from a connected device, which evidence a cloud platform can hand you for its layer, and which work stays with your team.

Short answer

The FDA regulates your device, and the cloud your device depends on is part of the system you submit. For a cyber device, section 524B requires a plan to address postmarket vulnerabilities, processes that provide reasonable assurance of cybersecurity, and a software bill of materials. A platform can supply audited controls and ready documentation for its layer. Your threat model, your risk assessment and your submission stay with you. Platforms are not FDA cleared or approved. Your device is the regulated product.

What section 524B requires

Section 524B of the FD&C Act applies to premarket submissions for cyber devices. A device is a cyber device if it meets all three criteria:

The sponsor must then provide three things:

  1. A plan to monitor, identify and address postmarket vulnerabilities and exploits in a reasonable time, including coordinated vulnerability disclosure.
  2. Processes that provide reasonable assurance the device and related systems are cybersecure, including patches on a reasonably justified regular cycle and out-of-cycle patches as soon as possible for critical vulnerabilities.
  3. A software bill of materials, including commercial, open-source and off-the-shelf components.

FDA's guidance, Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions, explains what reviewers expect. FDA states that manufacturers are responsible for identifying cybersecurity risks in their devices, including risks tied to cloud infrastructure. The guidance has been revised since 2023, so check the current version on fda.gov before you submit.

What FDA expects, and where the cloud layer fits

Submission elementWhat the platform layer can supplyWhat stays with you
Threat modelPlatform architecture and data flows for the cloud layerThe device-level threat model covering the device, its connection and your applications
Cybersecurity risk assessment with a scoring methodThe platform cybersecurity risk assessment (BioT CS-002)Your device risk assessment, with platform risks referenced
SBOM with NTIA minimum elements, support level and end-of-support datesThe platform SBOM, issued with every platform releaseThe SBOM for your device software and any code you deploy on the platform
Assessment of known vulnerabilities, including CISA's Known Exploited Vulnerabilities catalogStatic analysis and software composition analysis in BioT's CI/CD pipeline, with the issue detailThe same for your components
Architecture views: global system, multi-patient harm, updateability and patchability, security use casesPlatform architecture documentationYour system-level views
Penetration testing, with tester independence, scope and methodsIndependent penetration testing of the platformTesting of your device and your applications
Cybersecurity management plan: monitoring, patch timelines, coordinated disclosureThe platform cybersecurity management plan (BioT CS-001)Your device plan and your customer communication
Security labelingPlatform facts to referenceYour labeling

The platform documents, mapped to eSTAR

BioT maps its platform documentation to FDA's eSTAR template, version 7.0, so your regulatory team can reference each document by ID. eSTAR version 6.2 was retired on August 3, 2026.

DocumentIDUse in your submission
Software Requirements SpecificationSRS-0001Software documentation
Software Design DescriptionSDD-0001Software documentation
Software life cycle procedureQDC-00013Software development process evidence
Risk tableQRM-0003Software risk management
Test description and test reportSTD-001, STR-001Verification and validation
Software version descriptionSVD-0001Release identification
Cybersecurity Management PlanCS-001Cybersecurity management plan and patch timelines
Cybersecurity Risk AssessmentCS-002Cybersecurity risk assessment
Cybersecurity ReportCS-003Cybersecurity controls, architecture and testing
Software bill of materials, off-the-shelf software and vulnerability detailSBOM, OTS-001SBOM, off-the-shelf software and known vulnerabilities

Human factors content and every device-specific section remain yours. The full mapping, with eSTAR section numbers, is in the eSTAR submission support documentation.

IEC 62304 and the cloud layer

IEC 62304 governs the life cycle of your medical device software. You classify each software item by the harm it could contribute to: Class A, no injury or damage to health is possible; Class B, non-serious injury is possible; Class C, death or serious injury is possible. The class sets how much process evidence you need. Third-party software, known as software of unknown provenance (SOUP), needs defined requirements, a risk analysis and a rationale for its use.

BioT maintains an IEC 62304-compliant Design History File for the platform. It contains the software development plan, risk analysis, requirements and their review, design and its review, test plan, test description, test readiness review, test report, off-the-shelf software validation, full traceability, the software version description, the cybersecurity report and the SBOM. BioT documents the file as prepared to support software up to Class C.

The file is stand-alone. You can reference it, integrate it into your own Design History File, or extend it with your device requirements, risks and tests. Your software classification remains your decision.

When the platform releases a new version

Assess each platform release under your own change control and update your Design History File and ISO 14971 risk file where needed. BioT supplies release notes, an updated Design History File and an updated SBOM. Decide with your regulatory advisors which changes affect your submission.

Your own code on the platform

Custom code you deploy on the platform is your software. Develop it under your IEC 62304 procedure, keep its SBOM, and scan its repositories in your CI pipeline. It is not part of the platform SBOM.

Platform-layer controls, in detail

Certifications and attestationsHITRUST r2. SOC 2 Type II covering the Security, Availability, Confidentiality and Privacy criteria. ISO 27001 and ISO 27799 for the information security management system. ISO 13485 for the quality management system. Reports are shared under NDA.
EncryptionAES-256 at rest through AWS KMS. HTTPS only in transit, TLS 1.2 or higher. Devices connect over MQTT through AWS IoT Core with mutual X.509 certificate authentication.
NetworkAWS WAF inspects user and API traffic.
Access controlTwo layers. API permissions decide which calls a user can make. Attribute-based access control then checks every request to decide which data the user can reach. PHI fields are flagged per field and are visible only to the patient and to caregivers in the same organization. The manufacturer sees non-PHI data across organizations.
AuthenticationMulti-factor authentication with a one-time code by email or SMS. Passwordless one-time-code login. Enforced password policy.
Audit loggingUser activity is logged and audited, and logs are protected from tampering, including by privileged users.
Vulnerability managementStatic analysis and software composition analysis run in the CI/CD pipeline. Independent penetration testing of the platform.
BackupsProduction backups kept daily for 30 days, weekly for 12 weeks, monthly for 12 months and yearly for 10 years.
Deployment and contractsA dedicated environment for production, either in a BioT-managed AWS account or in your own AWS account. Hosting in the US, EU and Asia. BAA and DPA in both models, and a published subprocessor list.

When a hospital or a reviewer asks whether the evidence covers your data path, ask which components and regions each certification and attestation covers, and map them to your architecture. That is the question that comes up most in security reviews.

What stays with the device maker

BioT customers have achieved FDA clearance and CE marking on the platform. Those clearances belong to the customers and their devices.

The cost of building this evidence yourself

BioT estimates the direct certification and documentation cost of a do-it-yourself medical device cloud at about $945,000 over three years:

ItemThree-year cost
HITRUST r2$341,000
SOC 2 Type II$201,000
Design History File$132,000
Vulnerability scanning$78,000
ISO 27001$55,000
SBOM$51,000
ISO 13485$45,000
Penetration testing$42,000
Total$945,000

Time matters as much as cost. SOC 2 Type II and HITRUST r2 each require 18 to 24 months of documented operating history before certification is granted. See the compliance page for the full comparison.

Questions device makers ask

Is BioT FDA cleared or approved?

No. Cloud platforms are not cleared or approved. Your device is the regulated product. BioT provides certified, audited infrastructure and documentation that you reference in your submission.

How much cybersecurity work does the platform take off our team?

The platform layer: its controls, its certifications and attestations, its SBOM, its vulnerability scanning and penetration testing, and its cybersecurity documentation. The device-level threat model, risk assessment and testing remain your work.

Does the platform SBOM include our custom code?

No. Code you write and deploy is your software. Keep its SBOM and scan it in your own pipeline.

What do we tell FDA when the platform updates?

Assess each release under your change control. Changes to functions that serve your intended use, such as alerts or algorithms, deserve the closest review. BioT supplies release notes, an updated Design History File and an updated SBOM. Confirm the reporting decision with your regulatory advisors.

Can we use the documents directly in eSTAR?

Yes, by reference. Each platform document has an ID mapped to eSTAR sections, and you can reference, integrate or extend it.

Related: compliance and certifications, cloud infrastructure for SaMD, and EHR integration.

Build it Right

Book a demo